Q-Flex Trifocal 640MY

Single piece, sterile, foldable acrylic aspheric hydrophilic posterior chamber intraocular lens (IOL) with UV blocker and blue light filter. The optic is trifocal. The IOL is non-preloaded, which has to be manually loaded into a compatible injector.
Standard Powers:
+8.0 D → +30.0 D (0.5 D steps)
Extreme Powers:
-
Addition (at the IOL plane):
+3.5 D near
+1.75 D intermediate
Material:
Copolymer of hydrophobic and hydrophilic monomers with 25% water content, UV blocker and blue light filter
RI:
1,46
Abbe No:
58
Optic Diameter:
6 mm
Overall Diameter:
11.0 mm (8.0 D → +15.0 D)
10.7 mm (+15.5 D →+35.0 D)
Optic Design:
Aspheric - Aberration Neutral
Biconvex (+8.0 D → +35.0 D)
Nominal A-constant:
118,9
Contact Angle:
-
Diffractive Zone:
3 mm diameter diffractive array on the anterior surface utilizing EPS technology
Haptic Design:
Posterior vaulting 4 closed loops with 0° angulation
PCO Prevention:
360° sharp edge
Shelf Life:
5 years after sterilization
Estimated Incision Size:
1,8 - 2,2 mm
Compatible Injector:
Compatibility chart available:
medicontur.com/professionals/compatibility
LOADING GUIDE VIDEOS


Downloads

IOL Portfolio Digital
December 2, 2025
Q-Flex Trifocal datasheet (640MY) Digital
May 30, 2025
Q-Flex Trifocal datasheet (640MY) Print
September 15, 2025

Other products:

Trifocal Bi-Flex Liberty
Single piece, sterile, foldable acrylic aspheric hydrophilic posterior chamber intraocular lens (IOL) with UV blocker and blue light filter. The optic is trifocal. The IOL is non-preloaded, which has to be manually loaded into a compatible injector.
More details
1stQ AddOn® Trifocal
Single-piece, sterile, foldable acrylic aspheric hydrophilic posterior chamber intraocular lens (IOL) with UV blocker. The optic is trifocal. The IOL is non-preloaded and has to be manually loaded into a compatible injector.
More details
Bi-Flex T Preloaded
Single piece, sterile, foldable acrylic aspheric hydrophilic posterior chamber intraocular lens (IOL) with UV blocker, optically with blue light filter. The optic is monofocal toric. The IOL is preloaded, which is part of an injection system, delivered in a special lens holder, that must be inserted into a dedicated injector before implantation.
More details

MEDICONTUR WORLDWIDE

Medicontur Medical Engineering Ltd Hungary
Medicontur CZ s.r.o. Sales office Czech republic & Slovakia
Compagnia Italiana Oftalmologica scarl (CIO) Italy
SIA Industrialas Technologijas Latvia
ABJ-Vision Sp. z o.o Poland
Almaty Medical Products LLP Kazahstan
Heritage Medical Technology Jordan
Vision Matrix United Kingdom
Meraas Medical & Scientific Equipment L.L.C Syrian Arab Republic